2.2.7 Transferring the Patient
WARNING!
To reduce the risk of injury or damage:
-When elevated a few inches off the surface of the stationary object (wheelchair, commode, or bed) and before moving
the patient, check again to make sure the sling is properly connected to the hooks of the hanger bar. If any attachments
are not properly in place, lower the patient back onto the stationary object (wheelchair, commode, or bed) and correct
the attachments.
-Wheelchair wheel locks MUST be in a locked position before lowering the patient into the wheelchair for transport.
-Before transferring, check that the wheelchair weight capacity can withstand the patient's weight.
-Be sure to check the sling attachments each time the sling is removed and replaced to ensure it is properly attached
before the patient is removed from the bed or chair.
-The mast pivot under the rubber boot must be tight to ensure safe use of the patient lift. The bolt must be checked at
least every six months in conjunction with periodic maintenance.
2.2.8 Maintenance
WARNING!
-Maintenance MUST be performed ONLY by qualified personnel.
-The pump is sealed at the factory. DO NOT attempt to open the pump or obtain local service as this will VOID the
warranty and might result in damage. Consult your dealer or contact Invacare for more information.
-After the first year of use, the hooks of the hanger bar and the mounting brackets of the boom should be inspected every
three months to determine the extent of wear. If these parts become worn, they must be replaced.
-Casters and axle bolts must be inspected every six months to check for tightness and wear. After the first 12 months of
operation, inspect the hanger bar and the eye of the boom to which it attaches for wear. If the metal is worn, the parts
MUST be replaced. Repeat this inspection every six months thereafter.
-Regular maintenance of patient lifts and accessories is necessary to assure proper operation.
-DO NOT overtighten the mounting hardware. This will damage the mounting brackets.
-After the first six months of operation, inspect all pivot points and fasteners for wear. If the metal is worn, the parts
MUST be replaced. Repeat this inspection every six months.
13
1078987-O
Invacare® Reliant™ 450/600
Summary of Contents for Reliant 450
Page 16: ...3 Product Labeling Fig 3 1 1078987 O 16 3 Product Labeling...
Page 37: ...Fig 7 3 1078987 O 37 7 Patient Lifting...
Page 61: ...Fig 11 1 61 1078987 O Invacare Reliant 450 600...
Page 71: ...Notes 1078987 O 71...
Page 72: ...Notes 72 1078987 O Invacare Reliant 450 600...
Page 86: ...3 tiquettes sur le produit Fig 3 1 1078987 O 86 3 tiquettes sur le produit...
Page 106: ...Fig 7 3 106 1078987 O Invacare Reliant 450 600...
Page 129: ...Fig 11 1 129 1078987 O Invacare Reliant 450 600...
Page 138: ...Remarques 1078987 O 138...
Page 139: ...Remarques 139 1078987 O Invacare Reliant 450 600...
Page 140: ...Remarques 1078987 O 140...
Page 154: ...3 Etiquetado de Producto Fig 3 1 1078987 O 154 3 Etiquetado de Producto...
Page 174: ...Fig 7 3 1078987 O 174 7 Levantando al Paciente...
Page 211: ...Notas 1078987 O 211...