Invacare® Platinum® Oxygen Concentrator
6.3.5 SensO
2
Oxygen Purity Indicator
This feature monitors the purity of oxygen generated
by the oxygen concentrator. If purity falls below factory
preset standards, indicator lights on the control panel will
illuminate.
6.3.6 Initial Startup of the Concentrator
Concentrator may be used during the initial warm-up
time (approximately 30 minutes) while waiting for
the O
2
purity to reach maximum.
When the unit is turned on the yellow light will come on
indicating the unit is operating and is in the initial warm-up
period. When the green light comes on the unit is in normal
operating mode (SYSTEM OK/O
2
greater than 85%). Once
in normal operating mode the lights will be controlled
depending on the oxygen concentration values.
6.3.7 Explanation of Oxygen Purity Indicator Lights
Symbol
O
2
Purity
Indicator
Light
(LED)
Explanation
A
O
2
SYSTEM
OKAY
O
2
over
85%
GREEN
light
No action is
needed. Normal
for operation.
B
O
2
between
73% and
85%
YELLOW
light
(Solid)
Immediately call
supplier. You may
continue to use
the concentrator
unless instructed
otherwise by
your supplier.
Be certain that
backup oxygen is
nearby.
26
1195054-B