
SECTION 6—MAINTENANCE
28
FIGURE 6.2
Preventive Maintenance Record Invacare Platinum
Oxygen Concentrator
MODEL NO. IR C
SERIAL NO.
PREVENTIVE MAINTENANCE RECORD
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
- EVERY
4,380
H
OURS
Check Oxygen Concentration
UNITS WITHOUT
SensO
2
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
Check Power Loss Alarm
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
Check Power Loss Alarm
*NOTE:
Refer to Preventive Maintenance section of Service Manual.
NOTE:
4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.
26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
Check Oxygen Concentration
UNITS WITH SensO
2
ON EACH INSPECTION
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
- EVERY
2
6
,280
HOU
RS OR 3 YEARS, WHICHEVER COMES FIRST