60126903-A
10
3.2 Display
This section of the manual provides additional information about the display.
A
= Colored Dot
Icon
Indicator
Light Color
O
2
Concentration
Description Actions/Info
WHITE
N/A
Start/Stop
Press the Start/Stop button to start or stop the unit.
WHITE
N/A
Mute
Press the Mute button to mute alerts and alarms.
WHITE
WITH BLUE
DOT
< 82%
Start-up
The start-up symbol will rotate clockwise when illuminated. Refer
to
5.4.7 Initial Start-up of the Concentrator, page 28
WHITE
WITH
GREEN DOT
>82%
System
OKAY
No action is needed. Normal for operation.
WHITE
WITH
YELLOW
DOT
50% - 82%
Running
Alarms
Continue to use the concentrator, unless the unit has shut down.
Inspect the device:
l
Check cannula for kinks and check cannula connection to
concentrator. Refer to
1.1 Connecting/Positioning the
l
Check humidifier bottle connection. Refer to
the Humidifier (If Prescribed), page 1
l
Check and clean filter. Refer to
Refer to
for information about
specific error codes.
If the Alert continues after inspection, call provider.
Be certain that backup oxygen is nearby.
WHITE
WITH RED
DOT
<50%
Shut Down
Alarms
Invacare® Platinum® 5NXG Oxygen Concentrator