Invacare® Perfecto₂™ V Oxygen Concentrator
O
2
between 73% to
85%
YELLOW Indicator Light
A. YELLOW Solid
B. YELLOW Flashing
Sensor
Failure. Call a qualified
technician.
SYSTEM FAILURE
O
2
Below 73%
RED Indicator Light
Refer to Troubleshooting.
1.2 Intended Use
The intended function and use of the Invacare Perfecto
2
V
Oxygen Concentrator is to provide supplemental oxygen to
patients with respiratory disorders, by separating nitrogen
from room air, by way of a molecular sieve. It is not
intended to sustain or support life.
DANGER!
Risk of Injury or Death
This product is to be used as an oxygen
supplement and is not intended to be
life-supporting or life-sustaining. ONLY use this
product if the patient is capable of spontaneous
breath, able to inhale and exhale without the use
of a machine.
– DO NOT use in parallel or series with other
oxygen concentrators or oxygen therapy
devices.
DANGER!
Risk of Injury or Death
Depending on their medical condition, patients
on flow rates greater than 5 L/min may be at an
increased risk for serious injury or death in the
event of failure.
– ALWAYS discuss this increased risk with your
health care supplier BEFORE using this product
if you are prescribed a flow rate greater than
5 L/min.
WARNING!
Risk of Injury or Damage
Use of this product outside of the intended use
and specifications has not been tested and may
lead to product damage, loss of product function,
or personal injury.
– DO NOT use this product in any way other than
described in the Specifications and Intended
Use sections of this manual.