10.2 Electromagnetic Compliance (EMC)
Guidance and manufacturer´s declaration—electromagnetic emission
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the Device
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group I
The Device uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage fluctuations
/ flicker emissions
IEC 61000-3-3
Complies
The Device is suitable for use in all establishments including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Guidance and manufacturer´s declaration—electromagnetic immunity
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the Device
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 2, 4, 8 kV contact
± 2, 4, 8, 15 kV air
± 2, 4, 8 kV contact
± 2, 4, 8, 15 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.