Invacare® Perfecto₂™ V Oxygen Concentrator
9 Troubleshooting
9.1 Troubleshooting
SYMPTOM
PROBABLE CAUSE
SOLUTION
Alarm:
Main Power Loss:
No lights illuminated.
Short Beeps, long pause
Concentrator not operating,
power switch On.
Beep...Beep...
1.
Power cord not plugged in.
2.
No power at outlet.
3.
Tripped circuit breaker.
1.
Insert plug into outlet.
2.
Inspect house circuit breakers or fuses. If problem
recurs, use a different outlet.
3.
Push/reset circuit breaker. If problem recurs, call
service supplier.
Alarm:
System Failure:
RED light illuminated.
Continuous
Concentrator not operating,
power switch On.
Beep...
1.
Unit overheating due to
blocked air intake.
2.
Insufficient power at outlet.
3.
Internal repairs required.
1.
Do one or both of the following:
a. Remove and clean cabinet filters.
b. Move oxygen concentrator at least 30.5 cm (12
in) away from walls, draperies or furniture.
2.
DO NOT use extension cords. Move to another
electrical outlet or circuit.
3.
Call service supplier.
34
1195056-B