2.3 Radio Frequency Interference
WARNING!
Risk of Injury or Damage
To reduce the risk of injury or product damage
from interference with wireless equipment:
– Keep concentrator at least 3.0 m (9.8 ft) away
from wireless communication equipment such
as wireless home network devices, mobile
phones, cordless phones and base stations,
walkie-talkies, etc.
WARNING!
– Use of accessories and cables other than those
specified or provided by Invacare could result
in increased electromagnetic emissions or
decreased electromagnetic immunity of this
equipment and result in improper operation.
This equipment has been tested and found to comply
with EMC limits specified by IEC/EN 60601-1-2 for Type
BF equipment. These limits are determined to provide
reasonable protection against electromagnetic interference
in a typical home healthcare environment. Use of this
equipment in an environment with higher levels of
interference could affect the accuracy of the oxygen
concentration being delivered. This equipment is not
intended for use in a professional healthcare environment.
Other devices may experience interference from even
the low levels of electromagnetic emissions permitted by
the above standards. To determine if the emissions from
the concentrator are causing the interference, turn the
concentrator Off. If the interference with the other device(s)
stops, then the concentrator is causing the interference. In
such rare cases, interference may be reduced or corrected
by one of the following measures:
•
Reposition, relocate, or increase the separation between
the equipment.
•
Connect the equipment into an outlet on a circuit
different from that to which the other device(s) is
connected.
For additional information, refer to
Electromagnetic Compliance (EMC), page 39
.
1195056-B
13