Invacare® Perfecto₂™ V Oxygen Concentrator
Electrical Fast
transient / burst
IEC 61000-4-4
± 2 kV mains, 100kHz
± 2 kV mains, 100kHz
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to ground
± 1 kV line(s) to line(s)
± 2 kV line(s) to ground
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips
IEC 61000-4-11
0%
U
T
; 1/2 cycle @ 0,
45, 90, 135, 180, 225,
270, and 315
and
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30 cycles
Single phase: at 0°
0%
U
T
; 1/2 cycle @ 0,
45, 90, 135, 180, 225,
270, and 315
and
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30 cycles
Single phase: at 0°
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Device requires continued operation during
power mains interruptions, it is recommended that
the Device be powered from an un-interruptible
power supply or a battery.
U
T
is the a. c. mains voltage prior to application
of the test level.
Voltage
Interruptions
IEC 61000-4-11
0%
U
T
; 250/300 cycle
0%
U
T
; 250/300 cycle
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Device requires continued operation during
power mains interruptions, it is recommended that
the Device be powered from an un-interruptible
power supply or a battery.
U
T
is the a. c. mains voltage prior to application
of the test level.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m, 50/60 Hz
30 A/m, 50/60 Hz
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
40
1195056-B