SYMPTOM
PROBABLE CAUSE
SOLUTION
Alarm:
Potential Obstruction Alert
GREEN light illuminated.
Rapid
Beep...Beep...
Beep...Beep
1.
Possible internal obstruction
in the oxygen path. Kinked
or blocked tubing, cannula or
humidifier.
2.
Flowmeter set at 0.5 L/min or
less.
1.
Inspect for kinks or blockages. Correct, clean or
replace item. Once corrected, turn power Off for 60
seconds and then turn power back On.
2.
Reset flowmeter to prescribed flowrate. If prescribed
flowrate is below 0.5 L/min, use the Pediatric
Flowmeter IRCPF16AW.
The use of the Pediatric Flowmeter will deactivate
the Potential Obstruction Alert.
Alarm:
System Failure:
YELLOW light Illuminated.
Short beeps, long pause
Beep...........Beep.............
1.
Low oxygen purity.
2.
Kinked or blocked tubing,
cannula or humidifier.
3.
Flowmeter set at 0.5 L/min or
less.
1.
Clean or Replace filters.
2.
Inspect for kinks or blockages. Correct, clean or
replace item. Once corrected, turn power Off for 60
seconds and then turn power back On.
3.
Check flowmeter is set to 1.0 L/min or more. Refer
to
.
GREEN light with YELLOW
Light Flashing.
1.
Internal repairs required.
1.
Call service supplier.