
Invacare® LiNX
Remove screw/hand screw
A
and backrest shroud
B
.
2.
Fig. 6-3
Remove saliva trap
C
from tube.
3.
Insert new saliva trap with
INLET
imprinting facing
towards input module.
6.3 Cleaning Sip-N-Puff
Risk of Damage to Input Module
Improper mounting or maintenance of Sip-N-Puff
control may cause damage to input module by
water or saliva.
– Mouthpiece and breath tube MUST be
completely dry before installation.
Cleaning at least twice a week is recommended.
1.
Fig. 6-4
Remove mouthpiece
A
and lipswitch
C
from gooseneck
B
.
2.
Remove breath tube from saliva trap. See
3.
Position catch can beneath breath tube to collect water
and rinse.
4.
Flush mouthpiece and breath tube with warm running
water.
5.
Rinse with oral rinse to disinfect.
6.
Let dry completely before installation.
7.
Install mouthpiece, lipswitch and breath tube.
118
1637423-H
Summary of Contents for LiNX DLX-REM400
Page 124: ...Notes...
Page 125: ...Notes...
Page 126: ...Notes...
Page 127: ...www invacarelinx com...