SECTION 22—TDX BATTERIES
Part No 1125031
181
WARNING
When installing batteries, ensure battery connector is securely engaged to the con-
troller connector - otherwise serious personal injury may result.
4. Connect the controller to the batteries at the rear of the wheelchair.
5. Wheelchairs with TRRO and TRBKTS Only - Reinstall the top controller shroud using
the two thumb screws (FIGURE 22.4 on page 177).
6. Reinstall the rear shroud and secure in place with the existing two thumb screws.
NOTE: New Batteries MUST be fully charged before using, otherwise the life of the batteries will
be reduced.
7. Tilt the seat forward. Refer to Tilting the Seat Assembly on page 81.
8. If necessary, charge the batteries. Refer to Charging Batteries on page 184.
Cleaning Battery Terminals
WARNING
Most batteries are not sold with instructions. However, warnings are frequently
noted on the cell caps. Read them carefully.
DO NOT allow the liquid in the battery to come in contact with skin, clothes or
other possessions. It is a form of acid and harmful or damaging burns may result.
Should the liquid touch your skin, wash the area IMMEDIATELY and thoroughly
with cool water. In serious cases or if eye contact is made, seek medical attention
IMMEDIATELY.
1. Examine battery terminals for corrosion.
2. Verify the plastic caps are in place over battery cell holes.
3. Clean terminals by using a battery cleaning tool, wire brush, or medium grade sand
paper.
NOTE: Upon completion, areas should be shiny, not dull.
4. Carefully dust off all metal particles.