7.37
Adjusting the Seating System Mounting
Position
For Formula CG Seating System - Applies to Formula CG
Tilt, Recline or Tilt/Recline seating systems only.
1.
Loosen, but DO NOT remove, the four hex screws
A
and
locknuts
B
securing the seat frame mounting bracket
C
to the
interface mounting bracket
D
.
2.
Perform one of the following to ensure the Formula CG seating
system is in the proper mounting position:
Some front rigging and seat depth combinations may not
allow for the 1-inch gap. In this situation, slide the seat
frame mounting brackets as far back as possible.
•
Wheelchairs with Swing Away Front Rigging - Slide the seat
frame mounting brackets as far as possible towards the
rear of the wheelchair. Leave 1 to 1½-inch of clearance
between the front riggings and the front casters in all caster
positions).
•
Wheelchairs with Manual or Power Center Mount Front
Rigging - Slide the seat frame mounting brackets as far
as possible towards the rear of the wheelchair Leave 1
to 1½-inch of clearance between the center mount front
riggings and the front shroud.
3.
Ensure the interface mounting brackets and the seat frame
mounting brackets are flush and square.
4.
Secure the seat frame mounting brackets to the interface
mounting plates. Torque the four hex bolts and locknuts to 13
ft-lbs ± 20%.
5.
Cycle the tilt and/or recline functions to verify wiring harnesses
DO NOT obstruct the path of the system. If they do, perform
one of the following:
•
Wires were damaged during inspection - Replace damaged
wires.
•
Wires were not damaged during inspection - Cut tie-wraps
and relocate wires to a location where they will not become
damaged.
1143155-R-01
83