Invacare 

Flotech®

 

Contour Visco

Maintenance

7

6.2  Cleaning and care

!

IMPORTANT

All cleaning agents and disinfectants used must be effective, 

compatible with one another and must protect the materials 

they are used to clean. 

For further information on decon-

tamination in Healthcare Environments, please refer to ‘The 

National Institute for Clinical Excellence’ guidelines on Infec-

tion Control www.nice.org.uk/CG

139

 and your local infection 

control policy.

!

IMPORTANT

The Platilon covers are not washable and need to be replaced 

in case of heavy soilage.

Cleaning black covers

(Removal of contaminants such as dust and organic matter)

1.  Remove cover for laundering.
2.  Launder the cover at the recommended t

emperature between 60 

and 80 °C using a diluted detergent solution (instructions on label).

!

IMPORTANT

Washing at higher temperatures may cause shrinkage.

Drying black covers

1.  Hang cover from a line or bar and drip dry in a clean environment.

!

IMPORTANT

 

Tumble drying only for the ingress residence cover as this 

ensures that the fluorocarbon finish remains effective.

 

Tumble dry setting must not exceed 80 °C.

 

Do not tumble dry for longer than 10 minutes.

 

Dry covers thoroughly before re-fitting to the Flotech 

product.

Disinfecting black covers

(Reducing the number of microorganisms)

Please contact your hygiene specialist in the event of contamination.

!

IMPORTANT

Ensure that any residual detergent has been removed prior to 

disinfection.

Light soilage

1.  Wipe down the cover with a 0.1% Chlorine Solution (1,000 ppm).
2.  Rinse the cover with clean water using a single use nonabrasive 

cloth.

3.  Dry the cover thoroughly.

Heavy soilage

Where the cover is badly soiled, we recommend cleaning with a dilute 

cleaning detergent at 80 °C in the washing machine.

i

Large spillages of blood should be absorbed and removed with 

paper towels first.

1.  Clean up all spillages of bodily fluids i.e. blood, urine, faeces, spu-

tum, wound exudater and all other bodily secretions as soon as 

possible using a 1% Chlorine Solution (10,000 ppm).

2.  Rinse the cover with clean water using a single use nonabrasive 

cloth, and dry thoroughly.

3.  Dry the cover thoroughly.

!

IMPORTANT

1% Chlorine Solution used on a regular basis can diminish the 

life of the cover if not rinsed and dried properly.

 

Do not use granules.

WARNING

 

Remove contaminated foams from use.

CAUTION

 

Keep clear of open heat sources.

!

IMPORTANT

 

Do not use phenols, alcohols, bleaches, or other abrasive 

materials.

Summary of Contents for Flo-tech Contour Visco

Page 1: ...Invacare Flo tech Contour Visco This manual MUST be given to the user of the product BEFORE using the product read this manual and save it for future reference EN Cushion User Manual...

Page 2: ...ymbols on the product 5 3 Description 5 3 1 Product description 5 4 Usage 6 4 1 Safety information 6 4 2 Using the cushion 6 5 Transport 6 5 1 Safety information 6 6 Maintenance 6 6 1 Inspection 6 6 2...

Page 3: ...ded CAUTION Indicates a hazardous situation that could result in minor or slight injury if it is not avoided IMPORTANT Indicates a hazardous situation that could result in damage to property if it is...

Page 4: ...expectancy of three years for these products provided they are used in strict accordance with the intended use as set out in this document and all maintenance and service requirements are met The est...

Page 5: ...s Drip cables stents and other foreign objects must not become entrapped between the user and the pressure reducing surface of the cushion IMPORTANT The information contained in this document is subje...

Page 6: ...ntal cover damage please ensure that sharp objects are not placed on the cover When using patient transfer aids care should be taken not to damage the Flotech cover All transfer aids should be checked...

Page 7: ...umble dry for longer than 10 minutes Dry covers thoroughly before re fitting to the Flotech product Disinfecting black covers Reducing the number of microorganisms Please contact your hygiene speciali...

Page 8: ...a protective cover box Store cushions in a flat state on clean dry off flooring free from sharp edges to avoid any possible damage Never store other items on top of a cushion Do not store cushions nex...

Page 9: ...ights used as an indication Local requirements certification available upon request from Invacare 8 2 Material Viscoelastic foam Polyurethane foam components HR foam Polyurethane foam components Black...

Page 10: ...Invacare Flotech Contour Visco Technical data 10 8 4 Certificates of conformity...

Page 11: ...Invacare Flotech Contour Visco Technical data 11...

Page 12: ...Invacare Flotech Contour Visco Technical data 12...

Page 13: ...3 0 2 47 42 12 24 contactfr invacare com www invacare fr Invacare distributeur Invacare UK Operations Limited Pencoed Technology Park Pencoed Bridgend CF35 5AQ United Kingdom Tel 44 0 1656 776222 Fax...

Reviews: