Invacare® AVIVA™ RX series
Component
Check
Remedy
Notes
Battery mounting
Check battery support
and mounting belts for
damage
Replace if necessary
Check batteries for
damage
Replace batteries if
necessary
See
.
Check battery voltage
Charge batteries
See user manual.
Batteries
Check contacts and
terminals
Clean contacts and
terminals
See
.
Remote, status LED
flashing
Evaluate error/flash
code
See
Fixings
Tighten fixings, replace
if necessary
Cables and connecting
plugs
Replace damaged
cables and tighten
connecting plugs, if
necessary
Replace joystick knob
if necessary
Joystick function
Replace remote if
necessary
Remote / power
module
Power supply
Replace damaged
cables and tighten
connecting plugs, if
necessary
Chair configuration
Check chair
configuration version
Update software
if newer version
available
See LiNX service manual.
Screws (every 6
months)
Check screws for tight
fit
Tighten screws if
necessary
6.7 Shrouds
6.7.1 Power Module Shrouds
Removing Right-Hand Shroud
1.
Loosen hand screws
A
on right-hand side.
2.
Swivel shroud
B
forwards. Remove shroud sidewards.
Fig. 6-1
Removing Left-Hand Shroud
•
4 mm Allen key
24
1654647-B