10.5.4 Removing the Batteries from the Wheelchair
CAUTION!
Risk of Damage
Cleaning or maintenance may cause damage to
carpeting or flooring.
-Place the wheelchair in a well ventilated area where
cleaning or maintenance can be performed without
risk of damage to carpeting or flooring.
l
10 mm Hex Socket
1. Place the wheelchair in a well ventilated area.
2. Verify the battery isolation switch is in the Off position.
Refer to
7.3 Using the Battery Isolation Switch, page 48
more information.
3. Remove the rear shroud assembly. Refer to
Removing/Installing the Rear Shroud Assembly , page 75
4. Disconnect the left motor cable
G
from the
power module
P
.
Fig. 10-5 Disconnecting the motor cable
5. Using the 10 mm socket, remove the two bolts
M
securing
the battery door assembly
N
to the rear of the battery box
O
.
80
60101916-A
Invacare® AVIVA™ FX
Summary of Contents for AVIVA FX
Page 106: ...Notes...
Page 107: ...Notes...