Invacare 1085140 Assembly, Installation And Operating Instructions Download Page 1

NOTE: Check all parts for shipping damage before
using. In case of damage, DO NOT use. Contact the
Dealer/Carrier for further instructions.

SAFETY SUMMARY

The following recommendations are made for the
safe replacement of the FLAP KIT on the Reliant Bath
Lift (RBL250).

GENERAL WARNINGS

DO NOT assemble or use this kit without first
reading and understanding this instruction
sheet. If you are unable to understand the
Warnings, Cautions and Instructions in this
sheet and in the Owner's Operating and
Maintenance Manual (Part No. 1081003),
provided with the Bath lift, contact a health-
care professional, dealer or technical per-
sonnel before attempting to install this kit,
otherwise, injury or damage may occur.

DESCRIPTION

QTY

Flaps

2

Screws

4

Nuts - Self Locking

4

Washers

4

Hinges

4

Pins

4

Circlip

8

Instruction Sheet

1

This kit contains the following:

REPLACING THE BATH LIFT FLAPS

Removing the Bath Lift Seat (FIGURES 1-7)

NOTE: Refer to the GENERAL WARNINGS in the
SAFTY SUMMARY of this Instruction Sheet.

1. Raise the Bath Lift seat by pressing the UP (

é

)

button on the pendant (FIGURE 1).

NOTE: The lift will automatically stop at the top and
bottom limits if UP/DOWN buttons are not released.

INSTALLATION WARNINGS

When installing or removing the Battery
pack ALWAYS hold the battery pack firmly,
otherwise injury or damage may result.

To prevent frame from tipping when rais-
ing or lowering, either suction the frame
to the floor or place one foot on the frame
base when removing/installing the seat.

FIGURE 2 - REMOVING THE BATTERY PACK

PULL UP TO

REMOVE

é

Battery Pack

Frame

Handgrip

FIGURE 3 - RELEASING THE SEAT

WA R N I N G

To prevent pinching injury between seat
back and motor shroud when removing/
installing the seat, DO NOT grasp seat
anywhere other than at the recess at
the top (FIGURE 3), the opening at the
bottom of the seat or the lip where the
battery pack was.

CAUTION

DO NOT remove the seat until the bath Lift
is raised to the upper limit. Otherwise dam-
age may occur.

3. To release the seat, press and hold the spring loaded

chair release button (FIGURE 3).

Spring Loaded

Chair Release

Button

Recess

Seat

4. While holding the spring loaded chair release button

with one hand, grasp the seat bottom firmly with your
other hand.

5. Lift the seat up until the slide block comes free of the

grooves on the bath lift frame to remove the seat from
the frame (FIGURE 4).

2. Remove the battery pack by grasping the handgrip

on battery pack and pulling upwards (FIGURE 2).

1

FIGURE 1 - RAISING THE BATH LIFT

Pendant

Up

Motor Shroud

Battery Pack

Reliant Bath Lift (RBL250) Flap Kit

  Assembly, Installation and Operating Instructions

Kit No. 1085140