background image

USER GUIDE

Digital trimmer-controlled 
Behind-the-Ear hearing instrument 

STAGE 73 BTE
STAGE 83 PBTE

Summary of Contents for STAGE 73 BTE

Page 1: ...USER GUIDE Digital trimmer controlled Behind the Ear hearing instrument STAGE 73 BTE STAGE 83 PBTE...

Page 2: ...s important instructions for proper use and care technical performance information and other general informa tion about your hearing system Your hearing instruments have been adjust ed to your particu...

Page 3: ...e At night switch off your instrument and open the battery door completely It allows moisture in your instrument to evaporate and will increase the instrument s life span off Volume control Your instr...

Page 4: ...gram button A little later the instrument will give one single beep if set in program 1 two beeps if set in program 2 three beeps if set in program 3 When you close the battery door and switch the in...

Page 5: ...in the Ear Inserting the instrument With the battery door opened in the off position take the ear mold between thumb and index finger and position its point in your ear canal Now slide the earmold all...

Page 6: ...s blue and Right is red indication STAGE 73 STAGE 83 Recognizing left and right instrument If you have two hearing instruments they may be tuned differently One for your left ear the other for your ri...

Page 7: ...hone be sure to test it with your hearing system before purchasing For additional guidance please consult your cell phone provider Using the Telephone While Wearing your Hearing System Close proximity...

Page 8: ...attery by sliding it out as shown in the drawing Using Assistive Listening Tele Loop Systems Many gathering places are equipped with assistive listening tele loop systems for example in schools theate...

Page 9: ...showering swimming in heavy rain or in a moist atmosphere such as a steam bath or sauna The replacement battery has to be of type 13 zinc air Remove the protective seal from the fresh battery and ins...

Page 10: ...e battery door open Keep your instrument in a dry place not in a bathroom or other humid place Alternatively you could store the instrument in a hearing aid dryer available from your hearing care prof...

Page 11: ...strument use Intended Use Generic air conduction hearing instruments are wearable sound amplifying devices intended to compensate for impaired hearing The fundamental operating principle of hearing in...

Page 12: ...th a physician The exercise of such a waiver is not in your best health interest and its use is strongly discouraged Notice 1 A hearing aid will not restore normal hearing and will not prevent or impr...

Page 13: ...a child An audiologist is qualified by training and experience to assist in the evaluation and rehabilitation of a child with hearing loss iWarning to Hearing Aid Dispensers A hearing aid dispenser sh...

Page 14: ...form you on other facilities such as an induction loop at home for radio or television Eight steps towards better hearing You need to get used to your new hearing instrument Sounds seem new and differ...

Page 15: ...lowing conversations in a group is often difficult because of the background noise Listen to the different voices Try to recognize them by timbre or rhythm and link each voice to a person Focus your a...

Page 16: ...d Is the plastic tube or the earmold clogged or broken Is you ear full of wax Solving small problems Your Interton instrument is a reliable one In case of minor problems or malfunctions you may be abl...

Page 17: ...logged or broken Did your instrument get moist Did you leave your hearing instrument switched on at night Is the battery old SYMPTOM CAUSE POSSIBLE REMEDY Replace it with a new one Clean it or use a n...

Page 18: ...duction 21 dB SPL Frequency range DIN 45605 100 6030 Hz Current drain in test mode 0 95 mA Technical Data standard BTE Models ISG373 V Technical Data Power BTE Models ISG383 V Data in accordance with...

Page 19: ...y at least equivalent to the original hearing instrument As a signatory to the United Nations Global Compact initiative Interton is committed to doing this in line with environment friendly best pract...

Page 20: ...u have received Model received Hearing care professional ISG373 V ISG383 V Serial number Right Serial number Left Symbols i WARNING points out a situation that could lead to serious injuries CAUTION i...

Page 21: ...40 Interton 8001 Bloomington Freeway Bloomington MN 55420 Tel 1 800 247 4741 www interton usa com 16900811 Rev A 2013 01 HANDED OUT BY...

Reviews: