7.5 Using the TEOAE module
7.5.1 Preparing for the test
Patient Instruction
Place the patient on a bed or comfortable chair or on an examination table if necessary. Small children may
feel more comfortable sitting on a parent's or nurse's lap. Show the probe to the patient and then explain the
following:
•
The aim of the test is to screen the hearing organ for its functionality
•
The tip of the probe will be inserted into the ear canal and has to make a good seal
•
Multiple tones will be heard during the test
•
No participation is expected from the patient
•
Coughing, moving, talking and swallowing will disturb OAE test results
Visual inspection of the ear canal
Check the external ear canal for wax with an otoscope and remove excessive wax to prevent the probe
opening from clogging which will inhibit testing. Excessive hairs may have to be cut.
If contraindications exist the patient should be inspected by ENT or medical specialist.
Probe performance is crucial to OAE test results. We recommend that you conduct a probe test at the
beginning of each day before starting to test on patients to ensure that the probe is functioning correctly.
Prepare the equipment
1. Turn on Eclipse by connection the USB to the PC.
2. Open the OtoAccess® or Noah database and enter new patient details.
3. Double click on the IA OAE sw. icon to launch the software and click on the OAE module tab TE.
4. Select the desired test protocol from the dropdown list.
5. Select the ear for testing.
Before conducting the OAE test, ensure that the probe tip is clean and free of wax and/or debris.
Test environment
always conduct the OAE test in a quiet test environment, high ambient background noise
will affect the OAE recording.
6. Select an ear tip size that will ensure a good seal in the ear canal.
7. Check the probe check status in the software to verify a good seal.
When the probe is detected to be
Out of ear
the color is red,
In ear
the color is green.
When
Blocked
or
Too Noisy
the colored bar is amber.
8. Results for TE response testing, OAE present right side and no OAE response left side.
Summary of Contents for Eclipse ABRIS
Page 1: ...Instructions for Use EN Eclipse Science made smarter D 0120588 B 2019 03...
Page 2: ......
Page 80: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 72...
Page 112: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 104...
Page 132: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 124...