8.3 Impedance check
Press the Imp. button
and turn the dial fully
clockwise.
Turn the dial slowly
counter clockwise.
The LED for each
electrode will change
from red to green as
the impedance is
found for each
electrode.
Press the Imp. button
to exit the impedance
mode before testing.
8.4 Loop back (LBK15) testing
The LBK15 unit is not valid for performing a functional check of ABRIS due to the nature of the ASSR
algorithm. For functional testing (detection of false passes), a patient must be connected with electrodes, in
the absence of a stimulus (disconnected transducer).