background image

InTemp® CX400 Series Temperature Logger Manual 

19996-P 

InTemp CX400 series loggers measure temperature in indoor monitoring applications. Designed 
to meet World Health Organization (WHO) and Centers for Disease Control and Prevention 
(CDC) guidelines, this logger is ideal for clinical applications, such as vaccine storage and 
pharmaceutical manufacturing. This Bluetooth® Low Energy-enabled logger is designed for 
wireless communication with a mobile device. Using the InTemp app, you can easily configure 
the logger with one of four preset profiles designed for ambient storage, clinical refrigerator, 
freezer, or refrigerator monitoring or you can set up a custom profile for other applications. 
You can also quickly perform daily logger checks, download reports, and monitor tripped 
alarms. Or, you can use InTempConnect® to configure and download CX series loggers via the 
CX5000 Gateway. The InTempVerify™ app is also available to easily download loggers and 
automatically upload reports to InTempConnect. Use the built-in LCD screen on the logger to 
check the current temperature, daily maximum or minimum temperatures, logging status, 
battery use, and more. Once data is uploaded to InTempConnect, you can track logger 
configurations and automatically upload logger data to build custom reports for further 
analysis. The CX402 model uses a 2- or 4-meter probe and is available with a 5, 15, or 30 mL 
glycol bottle (bottle holder included). It also contains an internal sensor for ambient 
temperature monitoring. The CX403 model is available with an internal sensor only. 

Specifications 

External Temperature Probe

 

Range 

-40° to 100°C (-40° to 212°F) 

Accuracy 

CX40x-Txxx and CX402-BxM: 
±1.0°C from -40° to -15°C (±1.8°F from -40° to 5°F) 
±0.5°C from -15° to 80°C (±0.9°F from 5° to 176°F) 
±0.7°C from 80° to 100°C (±1.3°F from 176° to 212°F) 
CX40x-VFCxxx: 
±1.0°C from -40° to -22°C (±1.8°F from -40° to -8°F) 
±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F) 
±1.0°C from 50° to 100°C (±1.8°F from 122° to 212°F) 

Resolution 

0.024°C at 25°C (0.04°F at 77°F) 

Drift 

<0.1°C (0.18°F) per year 

NIST Calibration 

CX40x-Txx and CX402-BxM: Single point NIST calibration, probe and 
logger body 
CX40x-VFCxxx: Single point NIST calibration, probe only 
CX403: Single point NIST calibration 

Cable Length 

2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable 

Probe Dimensions 

Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches) 
long, 3.18 mm (0.125 inches) diameter 

Ambient Temperature Sensor

 

Range 

-30° to 70°C (-22° to 158°F) 

Accuracy 

CX40x-Txxx and CX402-BxM: 
±0.5°C from -15° to 70°C (±0.9°F from 5° to 158°F) 
±1.0°C from -30° to -15°C (±1.8°F from -22° to 5°F) 
CX40x-VFCxxx: 
±1.0°C from -30° to -22°C (±1.8°F from -22° to -8°F) 
±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F) 
±1.0°C from 50° to 70°C (±1.8°F from 122° to 158°F) 

Resolution 

0.024°C at 25°C (0.04°F at 77°F) 

Drift 

<0.1°C (0.18°F) per year 

Logger 

 

Radio Power 

1 mW (0 dBm) 

Transmission Range 

Approximately 30.5 m (100 ft) line-of-sight 

Wireless Data Standard 

Bluetooth Low Energy (Bluetooth Smart) 

Logger Operating Range 

-30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing) 

Logging Rate 

1 second to 18 hours 

InTemp CX400 Temperature 
Loggers 

Models: 

 

CX402-T205 & CX402-VFC205, 
2 meter probe and 5 mL glycol 
bottle 

 

CX402-T215 & CX402-VFC215, 
2 meter probe and 15 mL 
glycol bottle 

 

CX402-T230 & CX402-VFC230, 
2 meter probe and 30 mL 
glycol bottle 

 

CX402-T405 & CX402-VFC405, 
4 meter probe and 5 mL glycol 
bottle 

 

CX402-T415 & CX402-VFC415, 
4 meter probe and 15 mL 
glycol bottle 

 

CX402-T430 & CX402-VFC430 
4 meter probe and 30 mL 
glycol bottle 

 

CX402-T2M, CX402-B2M, & 
CX402-VFC2M, 2 meter probe 

 

CX402-T4M, CX402-B4M, & 
CX402-VFC4M, 4 meter probe 

 

CX403 

Included Items: 

 

Double-sided tape (for models 
with glycol bottles) 

 

Two AAA 1.5 V alkaline 
batteries 

 

Battery door and screw 

 

NIST Certificate of Calibration 

Required Items: 

 

InTemp app 

 

Device with iOS or Android™ 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for CX400 Series

Page 1: ...40 to 22 C 1 8 F from 40 to 8 F 0 5 C from 22 to 50 C 0 9 F from 8 to 122 F 1 0 C from 50 to 100 C 1 8 F from 122 to 212 F Resolution 0 024 C at 25 C 0 04 F at 77 F Drift 0 1 C 0 18 F per year NIST Calibration CX40x Txx and CX402 BxM Single point NIST calibration probe and logger body CX40x VFCxxx Single point NIST calibration probe only CX403 Single point NIST calibration Cable Length 2 or 4 mete...

Page 2: ...he logger for measuring temperature CX402 models only Audible Alarm Speaker This is the speaker for the audible alarm that beeps when an alarm is tripped or the external probe is removed if applicable Alarm LED This LED blinks every 5 seconds when an alarm is tripped or the external probe is removed if applicable See Logger Alarms LCD This screen shows the latest temperature reading and other stat...

Page 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 5: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 6: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 7: ...llowing two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate this equipment Industry Canada Statements This device complies with...

Reviews: