LSU220A : Load Support Unit
Operators Manual
3 INSTALLATION
3.1
POSITIONING THE EQUIPMENT
The LSU220A should be mounted on a level surface such as a workbench or sturdy shelf which
can support a weight of at least 30Kg.
The unit should be sited so that at least 10cm of airspace is available all round the left and top
of the unit and at least 20cm on the right hand side to allow free movement of air required for
cooling.
10 cm
20 cm
10 cm
1
2
BACO
VOLTAGE
CURRENT
OUTPUT STATE
OFF
OUTPUT TRIM
POWER
LSU220A
LOAD SUPPORT UNIT
ON
LOAD
RUNNING
STOPPED
OUTPUT 27V 220A MAXIMUM
Direction of Airflow
The LSU220A is supplied fitted with a three phase L1 + L2 + L3 + N + E plug on 3m of lead.
200.00 mm
360.00 mm
Total Length 3.0m Approx
P
lu
g
3P
+
N
+
E
IP
44
The load is connected via an Anderson 350A connector fitted to the chassis. A 3m lead is
supplied fitted at each end with mating Anderson 350A plugs. The cable has a cross section of
50mm2 per conductor. While operating the unit at very high currents the leads should be laid
out in such a manner as to aid cooling.