5 CALIBRATION
The LSU220A has two displays which show Voltage and Current. These are electronic
displays which will require regular checking for their accuracy. Intelligent Charging
recommends that the calibration check procedure is followed annually.
To calibrate the unit the following equipment will be required.
1.Calibrated voltmeter capable of reading 24V to one decimal place.
2.Calibrated ammeter capable of reading up to 200A to one significant digit, or a
calibrated ammeter shunt and millivoltmeter combination.
The equipment will need to be connected as shown in the following figures.
VOLTAGE MEASUREMENT
Doc: DWG1063-08-R2 LSU220A Operators Manual.odt
Page 10 of 16
Copyright Material of Intelligent Charging Limited © 2018
Printed On : 17/07/19
1
2
BACO
VOLTAGE
CURRENT
OUTPUT STATE
OFF
OUTPUT TRIM
POWER
LSU220A
LOAD SUPPORT UNIT
ON
LOAD
RUNNING STOPPED
OUTPUT 27V 220A MAXIMUM
1705 TRUE RMS PROGRAMMBLE MULTIMETER
+VE
COMMON
2A
10A
0.00V
0.00V