iQue
®
Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Clog Prevention and Management
Rinsing During a Plate
It is possible to configure a rinse at regular intervals while sampling wells on a plate. To configure a rinse, follow
these steps:
1.
In the Rinse section of the Protocol tab, check the box to
Enable Probe Rinse
and set the amount of wells
after which a rinse should occur. For example, to run a rinse after every well, choose
After every 1 well(s)
.
2.
Click the
Add
button. In the Add Rinse dialog, make sure the Rinse Station is set to
Buffer
and adjust the
Rinse Time (in milliseconds) as desired.
3.
Clicking
OK
will close the dialog. The Rinse Station and Time(ms) will now be displayed in the table.