Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Fluidics Maintenance
Changing the Detector Peristaltic Tubes
1.
Locate the sheath and waste pumps on the top of the detector.
Sheath Pump
Waste Pump
2.
Disconnect the Luer locks to the tubing on the outside of the peristaltic pumps by unscrewing them. The
blue tubing is connected to the sheath pump and the red tubing to the waste pump.