Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Fluidics Maintenance
Removing the Optics Cover
Prior to performing the first two steps of the fluidics maintenance procedure, the optics cover must be removed from
the detector. It is not necessary to shut down the detector before doing this as the optics cover contains an interlock
that automatically keeps the system from running when the cover is off.
1.
Open the lid of the iQue to expose the detector area. Locate the optics cover.
Optics Cover
2.
Using the 3mm hex key that came with the system, loosen the two captive screws that secure the optics
cover.
3.
Gently lift the optics cover off the detector area.