Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Recommended Maintenance Schedule
Recommended Maintenance
Schedule
Overview
iQue Maintenance involves several procedures that need to be performed regularly. What follows are guidelines for
typical usage scenarios.
Moderate Usage
is defined as less than 15 hours of run time per week.
Heavy Usage
is
defined as more than 15 hours of run time per week. These guidelines can be adjusted to account for the specific
conditions in any given laboratory.
Typical Usage Scenarios
Procedure
Moderate Usage
Heavy Usage
Daily QC
Daily at startup
Daily at startup
Daily Clean
Daily at shutdown
Daily at startup
Daily at shutdown
Quick Clean
As needed
As needed
Hydrocycle Clean
As needed
As needed
Change Probe and Tubing
(after 15 hours of run time)
Weekly
Every other day
Long Clean
Weekly
Every other day
Empty and Decontaminate All Bottles
Monthly
Every two weeks
Monthly Clean
Monthly
Every two weeks
Fluidics Maintenance Kit
Bi-Monthly
Monthly
Working with Samples that are Prone to Aggregation
Some samples can present special problems with clogging due to their tendency to aggregate. When working with
such samples the following additions can be made to the standard maintenance schedules:
•
Perform
daily after startup
•
Perform
daily prior to shutdown
See also,
Clog Prevention and Management
.
Summary of Contents for iQue SCREENER
Page 1: ...iQue SCREENER hardware manual October 2014...
Page 2: ......
Page 11: ...5 iQue Hardware Manual Part Number 10554 Rev C Introduction Declaration of Conformity...
Page 12: ...6 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Introduction...
Page 22: ...16 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Safety...
Page 64: ...iQue Hardware Manual iQue Maintenance 58 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 110: ......