iQue
®
Hardware Manual: Part Number 10554 Rev C., Safety
Operational Precautions
Ambient Temperature and Humidity
Operating temperature is between 15°C and 30°C (59°F and 86°F) and < 80% relative humidity.
Area of Use
•
The iQue Screening System is designed to fit on a laboratory bench top.
•
Provide a safety perimeter of 15 cm (6 inches) around the instrument and computer to allow proper
ventilation and to protect the instrument and computer from accidental liquid spills.
•
Take special care while handling fluids around the instrument.
•
Do not allow uncontained fluids to enter the interior of the instrument or computer.
•
Immediately clean up spills.
•
Do not place anything on top of the flow cytometer, including sample tubes and racks.
•
Before manually cleaning the system, turn off the instrument and unplug the power cord.
Electrical Precautions
Electrical devices pose the risk of an electric shock. To reduce the risk of an electric shock, do not open
the back or side service panels while the system is plugged in. Allow only authorized personnel to remove
any other panels from this device.
Always use the provided power cords, power supplies, and cables. Do not abuse the cords. Never use the cords to
pull the plug from an outlet. Keep cords away from heat, oil, or sharp edges. Damaged cords increase the risk of
electric shock.
CAUTION:
The instrument must be plugged into an IntelliCyt approved uninterruptable power supply (UPS). The UPS
must be plugged into a standard, grounded, or earthed mains electrical outlet, conforming to local codes. Non-
grounded or non-earthed mains outlet adaptors must not be used.
Summary of Contents for iQue SCREENER
Page 1: ...iQue SCREENER hardware manual October 2014...
Page 2: ......
Page 11: ...5 iQue Hardware Manual Part Number 10554 Rev C Introduction Declaration of Conformity...
Page 12: ...6 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Introduction...
Page 22: ...16 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Safety...
Page 64: ...iQue Hardware Manual iQue Maintenance 58 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 110: ......