Hardware Manual: Part Number 10554 Rev C., Special Topics: Re-Calibrating Deck Positions
Overview
iQue instruments are calibrated during their initial installation. But situations may arise where various points on the
deck need to be recalibrated. This procedure describes how to properly re-calibrate all the deck positions on an iQue.
NOTE
Deck positions should not normally need to be re-calibrated in the field.
Before performing deck position calibration, contact IntelliCyt Tech Support for
assistance to be sure that the problem is not a result of another issue such as
an improperly calibrated Plate Model or an improperly installed probe.
Required Materials
The iQue Screener comes with a special Calibration Tool which is required for recalibrating deck positions. One side
of the tool is for calibrating the well positions used for sampling a plate. The other is side is used for calibrating fluid
station positions.
Calibration Tool – Plate Calibration Side
Re-Calibrating Deck Positions
Summary of Contents for iQue SCREENER
Page 1: ...iQue SCREENER hardware manual October 2014...
Page 2: ......
Page 11: ...5 iQue Hardware Manual Part Number 10554 Rev C Introduction Declaration of Conformity...
Page 12: ...6 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Introduction...
Page 22: ...16 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Safety...
Page 64: ...iQue Hardware Manual iQue Maintenance 58 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 110: ......