The information and procedures described in this manual are intended to assist medical professionals in
the safe and effective use, care, cleaning, and long-term maintenance of INSURGICAL rechargeable batteries.
The INSURGICAL battery Cat. # 3002022 is intended for use with the INSURGICAL hand pieces only. The batteries must
be charged with the INSURGICAL MAXLiFe™ Battery Charger, Cat. # 3004022 ONLY.
INSURGICAL hand-held devices are provided STERILE. Contents of tray are sterile unless opened or damaged.
If tray or lid is opened, damaged, or if USE BY Date is exceeded, DO NOT USE. SINGLE-USE. DO NOT RESTERILIZE.
1) INSURGICAL batteries are for use by properly trained personnel who are completely familiar with
the required surgical techniques and instructions for use of the equipment.
2) Use of eye protection is required while operating equipment.
3) To assure the safety of the patient and operating room personnel, use only INSURGICAL rechargeable
batteries with INSURGICAL hand-held devices.
4) Do not expose battery to fire or incineration.
5) Do not allow battery contacts to contact metal objects. Contact with metal objects may result
in electrical shock, burn, or injury to user.
6) Do not disassemble or modify battery.
7) Do not incinerate or heat the battery.
8) Do not use damaged battery.
9) Avoid contact with leaking battery. In the event of eye contact with electrolyte, wash with copious
amounts of fresh water and seek medical attention immediately.
10) Do not sterilize battery.
With a cloth, wipe down the battery with a neutral enzymatic or neutral detergent solution before and after
each use. CAUTION: Do not use solvents to disinfect the batteries. Battery poles must not contact water or solvents:
Danger of short circuiting, causing burn or injury to user!
Storage Temperatur:
Operating Conditions:
Batteries can be stored separately in a dry cool place or left in charger that is powered on.
Charging Power: 19.54VDC
Output Power: 13.2VDC
Conforms to: UL1642, UL2054, EN 60601, UN38.3
1) Batteries contain Lithium Iron Phosphate (LiFePo4), which can be recycled or disposed of as municipal
waste. Battery should be fully discharged prior to recycling or disposal. Dispose or recycle in accordance
with your local government regulations.
2) Always use fully charged batteries to ensure maximum run time.
3) Battery charge cycle is 2.5 hours from full discharge to full charge. Typical charge time will vary depending
on amount of charge present when placed on charger.
INTRODUCTION
INTENDED USE
WARNINGS
CLEANING
ENVIRONMENTAL
SPECIFICATIONS
OPERATIONAL SPECIFICATIONS
/CERTIFICATIONS