7.4.2
Replacing the ionization chamber and sensor activation aid
Due to contamination remove the
ignition aid with the removing tool.
We recommend to rub the inside walls of the measuring chamber up to the groove for the retaining ring to a
bright finish using a polishing cloth.
CAUTION!
•
The sealing surfaces must only be worked
concentrically.
•
Do not bend the anode.
Insert the new ignition aid into the mounting
tool with the jagged side downwards
…
…
and slide it onto the anode until
the stop position is reached.