![Instron AVE2 Operator'S Manual Download Page 57](http://html1.mh-extra.com/html/instron/ave2/ave2_operators-manual_2070122057.webp)
57
Set up the load string
Product Support:
Set up the load string
Make sure the orientation is correct with respect to the camera on the video
extensometer. You will probably need to rotate the load cell mounting, any adapters, and
the base adapter so that the calibration bar and the specimen are facing the camera.
Caution
As far as possible, keep the extensometer away from exposure to conductive dust
or debris, from breaking carbon fiber specimens, for example.
Conductive dust or debris may cause a malfunction in the extensometer electronics. If
this is likely to be a problem, you are advised to surround the load string with a
transparent protective shield, separating it from the extensometer.
Set up for AverEdge32™ configuration
The AverEdge32™ configuration requires some additional setup to check that the
illumination is sufficient but not too bright.
1. Swing the back screen so that it faces towards the load string until it impacts on the
mechanical stop.
2. Install the set of grips that will be used for testing and a typical specimen.
3. In the AVE2 tab in the Transducer dialog, enable the Show saturated pixels option.
4. View the live image of the specimen and check for any saturated pixels, which display
as red dots in the image. If there are too many red dots to display on the image, a
warning message displays instead.
Lens focal length (mm)
Nominal field of view (mm)
Dot-based method
Edge-based method
35
130
not applicable
16 (standard)
310
205
9
560
not applicable
6
840
not applicable
Summary of Contents for AVE2
Page 1: ...Instron Video Extensometer AVE2 and SVE2 Operator s Guide M16 16905 EN Revision D...
Page 8: ...Preliminary Pages 8 M16 16905 EN...
Page 12: ...Preliminary Pages 12 M16 16905 EN...
Page 16: ...Chapter Introduction 16 M16 16905 EN...
Page 24: ...Chapter Specifications 24 M16 16905 EN...
Page 34: ...Chapter Installation 34 M16 16905 EN Figure 9 AVE2 on a 5940 single column testing machine A...
Page 46: ...Chapter Installation 46 M16 16905 EN Figure 21 AVE2 on an 8801 series testing machine...
Page 50: ...Chapter Installation 50 M16 16905 EN...
Page 76: ...Chapter Set up other software not Bluehill 76 M16 16905 EN...
Page 84: ...Chapter Maintenance 84 M16 16905 EN...
Page 87: ......
Page 88: ...www instron com...