49
System Performance
Product Support:
System Performance
Table 14. 34FM-100 system performance
Parameter
Specifications
Testing type
Tension, compression, and through zero
operation.
Basic control mode
Closed loop position control
Load capacity - kN (lbf)
100 (22480)
Maximum speed - mm/min (in/min)
508 (20)
Minimum speed - mm/min (in/min)
0.005 (0.0002)
Maximum force at full speed - kN (lbf)
50 (11240)
Maximum speed at full force - mm/min (in/min)
254 (10)
Return speed - mm/min (in/min)
600 (23.6)
Crosshead speed accuracy
± 0.2% of set speed at steady state and no
load
Displacement (crosshead position) display accuracy
Under no load conditions, equal or less than ±
0.02 mm (0.0008 in) or ±
0.15
% of displayed
reading, whichever is greater.
Displacement (crosshead position) repeatability - mm
(in)
± 0.015 (0.0006)
Load measurement accuracy
± 0.5% of reading down to 1/200 of load cell
capacity.
Strain measurement accuracy
± 0.5% of reading down to 1/50 of full scale
with ASTM E83 class B or ISO 9513 class 0.5
extensometer
Crosshead position control resolution - µm
0.06
Summary of Contents for 34FM-100