![Instron 300DX Operating Instructions Manual Download Page 79](http://html1.mh-extra.com/html/instron/300dx/300dx_operating-instructions-manual_2070128079.webp)
79
Hydraulic power supply (HPS)
Product Support:
Hydraulic power supply (HPS)
The HPS should be maintained according to the
“Preventative maintenance schedule”
.
Check the oil cleanliness
The cleanliness of the oil in the reservoir should be checked at the frequency stated in the
. Check the appearance of the oil through the sight gauge; it
should be a clear amber color. If the oil appears dirty, it should be changed (see
Add oil to reservoir
It should only be necessary to add oil to the reservoir if an oil leak is evident. However, the oil level
should be checked at the frequency stated in the
“Preventative maintenance schedule”
Check the oil level through the sight gauge on the HPS. The oil should be to the top of the sight gauge
when the frame’s hydraulic cylinder is fully retracted. When it is necessary to add oil, perform the
following procedure.
Equipment and supplies required
•
A 4 mm hex key (supplied)
•
A clean funnel with a mesh filter
•
Clean towels, rags, or similar
•
Hydraulic fluid (amount required depends on how much is needed to sufficiently fill reservoir), for
fluid specifications see
Figure 19. Pinion.
Pinion for 300DX
Pinion for 600DX
1. Groove for setscrew
2. Square end for crank handle
1
2
1
2
In the United States and Canada, an oil sample test kit is available for purchase from Instron Services;
contact your local Instron Services department as directed on page
for more information about the
test kit.
All equipment and supplies must be supplied by the customer unless noted as “supplied”. For a list of
equipment that was supplied with the system, refer to
Summary of Contents for 300DX