31
Troubleshooting
Product Support:
Before you contact Instron Service, note the model and serial numbers of the test system
and make sure there is a telephone at the test site.
Table 6. Troubleshooting
Problem
Cause
Remedy
Specimen slips under load
Incorrect size or type of jaw face
Install the appropriate jaw face
for specimen size and type
Cyclic or compressive loads
Do not use wedge action grips
for cyclic or compressive testing
Not enough gripping area
Install specimen for complete
engagement with jaw faces
Inadequate gripping force
Tighten grip handles
Inadequate jaw face lubrication
Lubricate rear and bottom of the
jaw faces
Load string misalignment
Verify alignment of load frame,
load string components, and
specimen
Specimen surface irregularities
such as scale or uneven surface
Clean and true specimen
surfaces
Specimen too hard
Do not test specimens harder
than 45 Rockwell C
Specimen breaks at jaw face
Initial gripping force is too great
for specimen
Do not over tighten control nut
Jaw faces do not release
specimen or fail to retract
Jaw face serrations are bound to
the specimen
Lightly tap specimen to release
bond
Debris or contaminants are
obstructing jaw faces
Remove and clean jaw faces,
clean the grip body, lubricate
and reinstall jaw faces
Tensile load on specimen
Remove load
Inadequate jaw face lubrication
Lubricate rear and bottom of the
jaw faces
Summary of Contents for 2716 Series
Page 8: ...Preliminary Pages 8 M10 14402 EN...
Page 12: ...Chapter Introduction 12 M10 14402 EN Figure 1 Typical Wedge Grip Front...
Page 16: ...Chapter Introduction 16 M10 14402 EN...
Page 20: ...Chapter Specifications 20 M10 14402 EN...
Page 24: ...Chapter Installation 24 M10 14402 EN...
Page 28: ...Chapter Operation 28 M10 14402 EN...
Page 32: ...Chapter Maintenance 32 M10 14402 EN...
Page 35: ......