Ventway Sparrow
User Manual
Page 134 of 144
23.15. BATTERY TYPE ALARM
Step
Procedure Action
Expected Results
1
Disconnect AC power
and insert a
rechargeable battery.
Start up the Sparrow
ventilator.
Complete the patient
circuit test.
The software switches to the ventilation phase, weight
selection screen.
2
Set the simulator to
the patient weight of
60 kg.
Select 60 kg and then
SIMV ventilation
mode, Invasive/Non-
Invasive, and confirm
the patient
connection.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Inspect battery icon.
Battery icon is of rechargeable battery, disconnected from
power.
4
Reconnect AC power. Battery icon is of rechargeable battery, connected to power.
5
Replace the battery
to non-rechargeable.
Disconnect from
power.
Inspect the icon.
Battery icon is of regular battery, disconnected from power.
6
Reconnect the AC
power.
Battery icon is of regular battery, connected to power.