
Ventway Sparrow
User Manual
Page 132 of 144
23.13. POWER ALARM
Step
Procedure Action
Expected Results
1
Start up the Sparrow
ventilator.
Complete the patient
circuit test.
The software switches to the ventilation phase, weight
selection screen.
2
Set the simulator to
the patient weight of
60 kg. Select 60 kg
and then SIMV
ventilation mode,
Invasive/Non-
Invasive, and confirm
the patient
connection.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Disconnect the
ventilator from the
power.
The software indicates Power is disconnected.
4
Inspect the battery
level.
Battery level is indicated to the user.
5
Reconnect the
ventilator to power.
The software indicates Power is connected.
Summary of Contents for Ventway Sparrow
Page 27: ...Ventway Sparrow User Manual Page 27 of 144 5 5 PNEUMATIC SECTION THEORY OF OPERATION...
Page 84: ...Ventway Sparrow User Manual Page 84 of 144 VWSP 100MR Civil Model VWSP 900 MR Robust Model...
Page 91: ...Ventway Sparrow User Manual Page 91 of 144...
Page 94: ...Ventway Sparrow User Manual Page 94 of 144...
Page 95: ...Ventway Sparrow User Manual Page 95 of 144...