23.9. LEAK ALARM
Step
Procedure Action
Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation
phase, weight selection screen.
2
Set the simulator to the patient weight
of 60 kg. Select 60 kg and then SIMV
ventilation mode.
Mode is set. The blower starts.
The software switches to running
ventilation, in SIMV mode.
3
Slowly increase a leak in the ventilation. After a short pause, the software
displays: “Leak Alarm”.
4
Close the leak.
The alarm disappears.