INOR APAQ C130 TC User Instructions Download Page 1

MEASURE OF SUCCESS

USER INSTRUCTIONS

Programmable

2-wire Transmitter

for Thermocouple

The user instruction must be read prior to adjustment and/or installation. 

All information subject to change without notice.

© INOR 10/2021 - 4008975501 - MA APAQ C130 TC R01 en

GENERAL INFORMATION

APAQ C130

TC

 

is a low cost, non-isolated, easy-to-use 

2-wire transmitter

, for

 

thermocouple input,

 intended to be used in industrial environment.

The transmitter is configured from a mobile device via 

NFC

 

c

ommunication, by using the app INOR Connect.  Inor Connect is 

available 

for free download on App Store, Google Play Store and Huawei AppGallery.

C130

TC

MEASURE OF SUCCESS

DATA  (short form)

CONFIGURATION

Input 

2-wire connection

TC type B - Pt30Rh-Pt6Rh (IEC 60584)

 

0...+1820 °C

 / 

+32...

+

3308 °

F

TC type E - NiCr-CuNi (IEC 60584)

 

-270…+1000 °C / -454...+1832 °F

Input impedance

10 MΩ

Maximum wire loop resistance

 

Output

 

Internal or fixed

Sensor failure indication 

Upscale (≥21.0 mA) or downscale (≤3.6 mA)

NAMUR Compliance

 

Current limitations and failure currents acc. 

Adjustable filtering 

l

evel

 

Ambient temperature

Storage and operation 

-40 to +85 °C / -40 to +185 °F

Galvanic isolation

 

None

Power supply 

6 to 32 VDC

Accuracy

 

Mounting

 

DIN B-head or larger

Configuration

 

App INOR Connect via 

NFC™ 

Factory settings

Input

0 to 10

0

0 °C

Output

4 to 20 mA

Sensor error indication

Upscale

This product should not be mixed with other kind of scrap, after usage.

It should be handled as an electronic/electric device.

www.inor.com, www.inor.se

INOR Process AB

, PO Box 9125, SE-200 39 Malmö, Sweden, 

Phone: +46 40 312 560, Fax: +46 40 312 570, E-mail: [email protected]

INOR Transmitter OY

, Unikkotie 13, FI-01300 Vantaa, Finland, 

Phone:+358 10 421 7900, Fax: +358 10 421 7901, E-mail: [email protected]

INOR Transmitter GmbH

, Am See 24, D-47279 Duisburg, Germany,

Phone: +49-203 7382 762 0, Fax: +49-203 7382 762 2, E-mail: [email protected]

KROHNE Temperature Division INOR

, 55 Cherry Hill Drive,

Beverly, MA 01915, United States 

Phone: +1 978 826 6900, Fax: +1 978 535 1720, E-mail: [email protected]

1.

TC type J - Fe-CuNi (IEC 60584

)

TC type K - NiCr-NiAl (IEC 60584)

TC type N - NiCrSi-NiSi (IEC 60584)

 

T

C type R - Pt13Rh-Pt (IEC 60584)

TC type S - Pt10Rh-Pt (IEC 60584)

TC type T - Cu-CuNi (IEC 60584)

-210…+1200 °C / -346...+2192 °F

-

270…+1300 °C / -454...+2372 °F

-

270…+1300 °C / -454...+2372 °F

-

50...+1750 °C / -58...+3182 °F

-50…+1750 °C / -58...+3182 °F

-270…+400 °C / -454...+752 °F

Cold Junction Compensation

 

In-head transmitter (TC sensor): 5 kΩ

4-20 mA, temperature linear

Warm-up time

After a max. 20 minutes the accuracy 

specifications are reached

(due to the internal cold junction)

RoHS, China RoHS

Directive: 2011/65/EU and 2015/863/EU

 

Harmonized standard:

 

EN IEC 63000

 

China RoHS 2

TC type 

R, S, T

 

TC type E

, J, K

 

TC type N 

(-100...+1300

 °C 

)

Max. of 

±1 °C or ±0.2% of span 

Max. of 

±1 °C or ±0.2% of span

TC type N 

(-270...-100

 °C 

)

±2.0 °C / ±3.6 °F

2.

Launch the app by clicking on the App

 

icon or holding 

your mobile device

 

against the transmitter on the part 

of the 

device where

 

NFC 

is located

 

(only possible with 

Android).

 

Click on “Read Configuration” and hold

 

your

 

mobile 

devic

against the trans

mitter as

 

explained in the 

first section.

 

 

a.

Sensor type

b.

Measuring rang

e

c.

Upscale or downscale

 

sensor

 

error indication

d.

F

ilter level

e.

TAG-number

f.

Password

3

.

4

.

Type K

TC type B 

Max. of 

±2.0 °C or ±0.2% of span

to NAMUR NE 43

0.4 to 26 

sec.

See Data Sheet

Max. of 

±3.6 °F or ±0.2% of span

Max. of 

±1.8 °F or ±0.2% of span

Max. of 

±1.8 °F or ±0.2% of span

Once you have set the desired values in the app, you transfer it to the 

transmitter by clicking the transfer button and holding the mobile device 

against the transmitter until a green check box appears confirming that 

the transfer has been completed.

CJC accuracy 

Typical 

±1 °C 

±

1.8 

°F (max ±

3 °C

 / 

±5.4 °F

) within ambient temperature range

Before making a configuration of APAQ C130

TC

 y

ou need to do

 

following:

 

Make sure that you have a mobile device with NFC

 

communication 

activated.

 

Download the app INOR Connect to your mobile device.

Required versions: 

 

iOS:

 iOS 13 or later and

 

Android

:

 

 

Iphone 7 or later for NFC

 

Android 4.4 

or later

 

Configuration procedure:

In the app you can edit the following: 