19P03-DOC01E REV-3
38
Preventive maintenance
Checks to be conducted annually:
▪
Check that the equipment is still in good condition;
▪
(If applicable) Connect the device power cable;
▪
Leave the device connected for at least 48 consecutive hours;
▪
Verify that the nuts and bolts are tight;
▪
Check the wear of shoulder washers used to reduce friction of moving parts. Do not
lubricate. Replace if necessary;
▪
Check that no object or equipment hinders the movement of the moving parts of the
bed (wheels, structure, etc.);
▪
Check the wire condition of; the remote control, the power cord and bed cables. Replace
damaged ones;
▪
Verify correct operation of all functions of the device according to the instructions for
use provided in this manual (see section 3).
▪
Unplug the power cable and operate the actuators to verify the battery.
WARNING ABOUT PREVENTIVE MAINTENANCE
▪
Preventive maintenance is essential to maintain the reliability of the bed. A bed
which demonstrates abnormalities should not be used. Any abnormality must be
repaired immediately by an authorised and competent person. Failure to follow
these safety instructions for cleaning would jeopardise the integrity of the bed and
could lead to
SERIOUS INJURY
.
▪
Any replacement of an equivalent part not certified by Rotec International on the
device may result in
SERIOUS INJURY AND / OR SIGNIFICANT
MECHANICAL BREAKDOWN AND MAY LIMIT OR VOID THE
PRODUCT WARRANTY.
▪
Any break on a part giving access to electronic parts must trigger an immediate
lockdown of the device until the broken part is repaired/replaced. Not doing so
could lead to
SERIOUS INJURY BY ELECTROCUTION
.