Inappropriate installation and initial
operation
WARNING!
Risk of injury due to inappropriate installation
and initial operation!
Inappropriate installation and initial operation may
lead to severe damage to persons or loss of
property.
Therefore:
– Make sure that enough space is provided for
the installation before starting.
– Take care regarding sharp-edged components
– Keep the installation site tidy and clean!
Discarded components and tools may be a
source of accidents.
– Component parts must be installed
professionally.
– Secure component parts so that they cannot fall
off or tumble.
6.2 Location requirements for installation
6.2.1 Place of installation
The following requirements must be met for the installation of the
unit:
The equipment must be installed and operated inside a
building.
The unit must not be exposed to direct sunlight.
NOTICE!
The incubator shaker must be protected against
hot exhaust air of other equipment with cooling
systems.
Ambient temperature must lie within the following range:
Minimum: 1 °C
Maximum: 30 °C
Without cooling: not more than 5 °C below the minimum
incubation temperature
With cooling: not more than 20 °C above the minimum
incubation temperature
Protect the unit against mechanical damage (trolleys, being
kicked, etc.).
The installation site must be level, sufficiently stable and able to
bear loads.