NOTICE!
An optional control program is available for
simultaneous operation of the humidification and
the cooling system.
If the illumination is switched on, the maximum humidity may
not reach 85% rH under all conditions.
Fig. 34
The Direct Steam Humidification is mounted on the rear of the
housing.
A controlled pump feeds water dropwise into an evaporator
chamber via a feed hose. The steam is then fed directly into the
incubation chamber.
There are two types of water supply:
unpressurised from a water tank
pressurised to max. 0.3 bars from a pressurised pipeline
4.11 Option CO
2
gassing
The incubator shaker is equipped with a CO
2
gassing unit.
This comprises:
A controlled CO
2
gassing device
A transparent safety curtain inside the chamber (the gassing