Multitron Pro - Incubation Shaker
Maintenance
2013-02-19
Page
145
of
185
8.4 Replacing the fuse
Replace the fuse as follows:
Procedure
Fig. 100
1
.
Unlock the fuse link by pressing the snap lock.
2
.
Remove the fuse link.
3
.
Replace defect fuses.
4
.
Fit new fuse link.
5
.
Lock the fuse link by pressing until it clicks into place.
8.5 Cleaning the cooling system
The cooling fins of the condenser must kept free of dust to allow
optimum heat dissipation. This requires regular inspections and
cleaning.
CAUTION!
Risk of damaging the heat exchanger due to the
use of sharp objects.
The copper pipes of the heat exchanger may be
damaged by cleaning with sharp objects, thus
leading to the escape of coolant.
Therefore:
– Use only a vacuum cleaner to clean the fins and
copper pipes of the heat exchanger