7.20 Starting algae illumination
The illumination is switched on using the operating panel.
Procedure
1.
Press option key to select parameter illumination.
Parameter sign „Light“ lights up.
2.
Press SET key.
3.
Press ON/OFF key to switch parameter on.
Display shows ON.
4.
Set the light intensity with the rotary knob.
7.21 Operating the CO
2
gassing system
NOTICE!
To prevent excessive consumption of gas, the
door, drain hole and path-through (optional) of the
incubation shaker must be closed.
NOTICE!
The CO
2
gassing system only works correctly when
the temperature control system is switched on.