Multitron Pro - Incubation Shaker
Operation
2013-02-19
Page
127
of
185
7.12 Handling the flasks without pulling out the tray
The flasks can also be handled without having to pull out the tray.
The flasks are handled as follows:
Procedure
1
.
Interrupt the incubator shaker (see chapter
Interrupting
operation
)
NOTICE!
The flasks can be handled when the motor has
stopped and the door lock has disengaged
(metallic clicking noise).
2
.
Open the door to a maximum of 80°.
NOTICE!
The tray is disengaged when the door is opened by
more than 80°.
3
.
Handle the flasks.
4
.
Close the door.
The incubator shaker automatically starts operating again with the
preset parameters.
7.13 Loading the trays
CAUTION!
Risk of material damage due to excessive wear
resulting from uneven loading of trays.
Uneven loading of trays considerably shortens the
service life of the suspension mechanism and
bearings of the table.
Therefore:
– Always place the flasks symmetrically with
respect to the centre of the tray.
The unit must be switched on when loading the table so that the
safety lock on door will disengage and the door can be fully
opened.