Multitron Pro - Incubation Shaker
Operation
2013-02-19
Page
109
of
185
From 1 day 0 hours up to 9 days 23 hours
7.4.2 Continuous operation without timer
The timer is switched off.
The timer function symbol does not light up.
The shaker operates continuously with the preset parameters.
7.4.3 Setting the timer
The timer can be used to change over between two operating
modes. This can be a single changeover to continuous
changeover.
Time 1 and Time 2 can be set with the TIMER key.
In operating mode Time 1 all parameters are set, that are active in
Phase 1.
In operating mode Time 2 all parameters are set, that are active in
Phase 2.
NOTICE!
Do not activate the timer in Time 2. This starts the
programmed cycle.
CAUTION!
Using the «Sticky Stuff» adhesive matting,
there is a risk of loss of property due to
condensation caused by sudden drops in
temperature!
Do not program jumps in temperature with the
timer, while using «Sticky Stuff» adhesive matting.
This may lead to condensation and thus cause the
cultivation flasks to detach from the adhesive
matting!
Therefore:
– Never program temperature jumps with the
timer.