Multitron Pro - Incubation Shaker
Installation and initial operation
Page
100
of
185
2013-02-19
Fig. 78
3
.
Push the cooling return hose onto the male outlet nozzle.
4
.
Secure the cooling return hose with a hose clamp.
6.12 Connecting the Direct Steam Humidification
CAUTION!
Risk of material damage due to build up of lime
scale through use of tap water.
Lime scale may build up in the evaporator and the
incubation chamber through the use of tap water.
Therefore:
– Use only the recommended types of water for
humidification.
Before starting initial operation the Direct Steam Humidification, it
must be connected to an unpressurised water tank or to a
pressurized water supply.
The inlet pressure must not exceed 0.3 bars.
NOTICE!
Place the water tank on the floor to avoid damaging
the equipment and the electronics.
The water tank must fulfil the following conditions:
closed
capable of being sterilized
equipped with a filter for pressure compensation
sufficient filling for one cycle
located <2 m away from the incubator shaker
Altitude <1.60 m above floor, if not placeable on
the floor.