Use
r’s
Manual of OMNI Express Patient Monitor
85
037.007.0001 Rev.2
RECALL DATA
◼
RECALL DATA STORAGE
◼
RECALL DATA DISPLAYS
◼
RECALL OPERATION DESCRIPTIONS
RECALL DATA STORAGE
Recall Data in graphical or tabular format can be displayed on the screen, transferred to a
computer for analysis via the RS232 interface, or printed if a printer is installed.
The recall data for all parameters is the average of a 5-second sample of the data. Sixty
(60) hours of recall data is stored in a nonvolatile memory, and remain in storage when the
monitor is in Standby.
A new print of recall data is started each time the monitor is turned on. A recall data record
is defined as the data from one Power On event to the Standby power event. A date/time
annotation is included at the start of each new print (for up to eight patients) and the print
can be collated with the patient. Once the recall memory has stored 60 hours of data, the
oldest recall data will be overwritten by new data.
RECALL DATA DISPLAY
The Recall data are displayed in graphical or tabular format. The recall information in
graphical format for a selected parameter is shown as a line graph connecting each of the
points representing the stored 5-second average.
The information stored for each recall episode can include:
• Numeric vital signs for all the measurements monitored
• Waveforms for up to 12 measurements of alarm events for your choice
You can navigate through the recall database to view events retrospectively, and you can
document recalls on a recording or report marked with the patient name, patient ID, the
data and time.
RECALL OPERATION DESCRIPTION
1. You should enter the ID and name of a patient first for recall
After you power on the monitor, there will be a pop up on the screen to remind you to the
patient’s ID:
Figure 33: Window for Indication Information
The above window will be automatically closed after 10 seconds.
2. Press the
“Recall” soft-key to open the recall function for up to 8 patients
3. Select the patient
’s ID for recall