Manual of OMNI Express Patient Monitor
93
PERIODIC SAFETY CHECKS
It is recommended that the following checks be performed every 24 months.
• Inspect the equipment for mechanical and functional damage.
• Inspect the safety labels for legibility.
The Life expectancy of the OMNI Express Patient Monitor depends on how frequent and
how long it is used. For a properly maintained, its life expectancy is 5 years. For more
aggressive use models, life expectancy can be less. We recommend replacing the
monitor every 5 years.
CLEANING
To clean the OMNI Express, dampen a cloth with a commercial, nonabrasive cleaner and
wipe the exterior surfaces lightly. Do not allow any liquid to come in contact with the power
connector or switches. Do not allow any liquids to penetrate connectors or openings in the
instrument. For cables, sensors, and cuffs, follow the cleaning instructions in the
instruction manuals for these accessories.
[WARNING]:
DO NOT spray, pour, or spill liquid on OMNI Express, its
accessories, connectors, switches, or openings in the chassis. DO NOT submerge
the OMNI Express or its accessories in liquid or clean it with caustic or abrasive
cleaners.