Connector Description
Precision 2.4mm connectors are miniature, instrument-grade, air-dielectric connectors that operate mode-
free up to 50 GHz. They feature extremely low VSWR and insertion loss and are designed to non-
destructively mate with standard 2.4mm connectors. These connectors generally have a high-performance
support bead and comply with the proposed
IEEE Standard 287 for Precision Coaxial Connectors
.
Connector Care
Precision connectors must be handled carefully if accurate calibrations and measurements are to be
obtained. All connectors should be inspected prior to each use. For optimal measurement results, all
interfaces should be visually inspected under magnification and cleaned on a regular basis. Proper
connector contact pin depths should also be verified through regular inspections using a connector gage
kit to ensure that connectors on both calibration devices and devices under test (DUTs) have contact pin
depths within recommended tolerances.
Care should be used whenever aligning connectors. Tighten connector coupling nuts using an appropriate
torque wrench while holding the opposing connector with an open-end wrench.
When disconnecting devices, take care not to rock or bend any of the connections. Disconnect devices by
disengaging the coupling nuts and gently pulling the connectors apart in a straight line.
Always use protective covers on all connectors when devices are not in use.
Should a connector become damaged, it should be repaired before it is used again or replaced immediately.
A damaged connector can damage other mated connectors.